RayTek 54301 Thermometer User Manual


 
ProgrammingGuide
11.7MultipleUnits(MultidropMode,RS485)
Upto32unitscanbeconnectedwithinaRS485network,see section
5.6
InstallingofMultipleSensorsviaRS485onpage31.Todirecta
command to one unit among the 32 possible, it is necessary to
„address“acommand.Therefore,a3digitnumberissetpriorthe
command.The3digitnumberisdeterminedbetween001and032.
Exception:Broadcastcommand.
Ifacommandistransferred,starting
withthe3digitnumber000,all
units (with addresses from 001 to 032) co nnected will get this
commandwithouttosendananswer.
Note: A unit with the address 000 is a single unit and not in
multidropmode.
XA=024 willsetaddressto24(unitmustnotbe
inmultidropmode)
Changinganaddress:
(e.g.theaddressischangefrom24to17)
command answer
„017?E“ „017E0.950“
„017XA=024“ “017XA024”settingofanewaddress
„024?E“ „024E0.950“
Exampleforthebroadcastcommand:
command answer
“024?E” “024E0.950”
“000E=0.5” willbeexecutedfromallunits,noanswer
“024?E” “024E0.500”
“012?E” “012E0.500”
MID 75