RayTek 54301 Thermometer User Manual


 
ProgrammingGuide
XF factorydefaultvalueswillbeset
11.6.5LockMode
Theaccesstotheunitispossibleviaserialinterface(software)and
viathedirectuserinput(modebuttons,LCDdisplay).Itispossible
tolockthebuttons.Thisallowsaccesstheunitonlyviasoftware.
J=L directuserinputvia
modebuttonsdenied
Alternativelytheunitcanbeunlockedbypressingthe<Mode/Up>
buttonssimultaneouslyforthreeseconds.
11.6.6ModeSettingfortheDigitalInputFTC3
ThedigitalinputFTC3(seesection5.4.4
TriggerandHoldFunction
onpage28)canbeusedasfollows:
XN=T FTC3astrigger
XN=H FTC3withholdfunction
11.6.7ChangingtheSensingHeadCalibrationData
Ifitisnecessarytoexchangethesensinghead,youmustsetthe
calibrationdataforthenewsensinghead:
XZ=01234567FFFFFFFF accordingto
theheadcalibrationdata
ForMID/MICmodelsthelasteightnumbersarenotusedandmust
besetto„F“likeshownintheexampleabove.ForMIHmodelsand
specials(likeMTBorG5)allsixteennumbersareused.
11.6.8AmbientBackgroundTemperatureCompensation
Incaseofcompensatingtheambient
backgroundtemperature,the
followingmodesareavailable:
AC=0 nocompensation
AC=1 compensationwithaconstanttemperaturevalueset
withcommandA.
MID 73