Installation
5.4.4TriggerandHoldFunction
TheFTC3inputcanbeusedasexternaltriggerinconjunctionwith
thesoftwaretriggermodesetting“Trigger”or“Hold”.
External switch:
- contact relay,
- transistor,
- TTL gate, …
Figure17:WiringofFTC3asExternalInput
Trigger:AlogicallowsignalattheinputFTC3willresetthepeakor
valleyholdfunction.Aslongastheinputiskeptatlogicallowlevel
thesoftwarewilltransfertheactualobjecttemperaturestowardthe
output. At the next
logical high level, the hold function will be
restarted.
Temp
object temperature
output temperature
FTC3
Figure18:FTC3forResettingthePeakHoldFunction
Time
28 MID