54
MESSAGE WHAT IT MEANS
1) You may have high
glucose and have
tested in an
environment near
the low end of the
system’s operating
temperature
range(6–44°C/
43–111°F ).
2) There may be a
problem with the
test strip. For
example, it may have
been damaged or
moved during
testing.
3) The sample was
improperly applied.
ACTION
If you tested in a cool
environment, repeat
the test in a warmer
environment with a
new test strip.*
If you were testing in a
normal or warm
environment, repeat
the test with a new test
strip. (See pages 10-11
for test strip
information.)*
If you applied the blood
incorrectly, review
pages 30-31 on blood
application and repeat
the test with a new test
strip.*
*If the error message
appears again, contact
OneTouch
®
Customer
Care Line at
1 800 663-5521.
Error message that indicates one of the following conditions may be present:
DCO_06052203A_OTU_OB_CAN_EN 2/20/04 1:39 PM Page 54