Leica 106Z Microscope & Magnifier User Manual


 
15
3
1
2
4
5
6
Hg50W
Hg100W
Adjusting the mercury vapour lamp
This adjustment is important for obtaining a uniform light spot
and good-quality fluorescence.
While adjusting the mirror image, do not project light
on to the electrodes for long periods (risk of explosion
through overheating)
The two electrodes are imaged as an extension of the plane
of symmetry of the discharge arc, but are difficult to see.
Switch on the supply unit and wait for two or three minutes
• The100Wsupplyunitemitsawhistle
Swing out the UV-light excluder (page 21)
• WorkwithouttheUV-lightexcluderisonlypermittedforadju-
stment, and then only for short periods
Pull out the light stop (see page 16, MZ10 F and page 22,
Fluorescence module).
MZ10 F: Turn the filter set in the beam path (page 16).
Stereo-fluorescence module: Insert the excitation filter
(page 22).
Mark a cross on a piece of paper and place it in the middle
of the illuminated spot
Select the lowest magnification
Figs. a: Bringing the discharge arc
into focus
Figs. b: Positioning the discharge arc
relative to the cross
Figs. c: Focusing and positioning the
mirror image of the discharge arc
• Thedischargearcisvisibleintheilluminatedspotandlies
at about 45° to the lines of the cross (fig. a)
Look into the eyepieces and focus on the cross
Using your unaided eye, observe the discharge arc on the
paper and bring it into focus with the focusing knob (3) (fig. a)
Using the positioning knobs (1 and 2), displace the discharge
arc (fig. b)
Bring the mirror im age of the discharge arc into focus
with the knob (6) and use the knobs 4 and 5 to position it
symmetric ally relative to the original image (fig. c)
• Forthe50Wburner,thedischarge-arcimageandits
mirror image should touch
• Forthe100Wburner,thedischarge-arcimageanditsmirror
image should be superimposed.
Using the focusing knob (3), readjust the illuminated field
• Theilluminatedfieldshouldnowbelarge,circularand
as uniform as possible
Reposition the UV-light excluder correctly (page 21).