Leica 106Z Microscope & Magnifier User Manual


 
13
Using the focusing knob,
displace the collector:
• Duringthismovement,the
collector must not touch
the flexible power cable. If
necessary, bend the cable
away
Carefully close the lamp housing, making sure that
the dis connect plug engages the socket
Press in the disconnect plug slightly
Tighten the screws
Push the disconnect plug fully in to the stop
without turning it.
Lamp housing assemble
Using the Allen key, release
the hollow screw
Introduce the lamp housing
into the adapter of the
fluo res cence module and
tighten the hollow screw.
Stray-light protection
There are slots in the lamp housing through which UV light can
shine on the user's hands during manipulations beneath the
housing. The stray-light protection on the underside of the lamp
housing is designed to block this light.
Secure the stray-light
protection with two screws.
Removing the collector and the heat-protection filter
The collector and the heat-protection filter can be removed
for cleaning or if defective.
Open the lamp housing as described on page 12 and remove
the burner socket
Move the reflector fully to
the right-hand stop
Holdthecollectorandpull
out the focusing knob
Extract the collector
Slacken the two screws
and remove the heat-
protection filter from the
collector
Screw a new heat-
pro tection filter into
position
Replacing the collector
Pull out the focusing knob
Carefully replace the collector and move it to the left until
the focusing knob engages the guide groove
Test the displacement of the collector
Refit the burner socket (page 12) and close the lamp housing.