Leica 106Z Microscope & Magnifier User Manual


 
14
Assembly
Supply unit for lamp housing 106Z
with 50W mercury-vapour burner
1. On/off switch
2. Running-time meter
3. Selector for L1/L2 burners
4. Fuse
5. Selectorfor50Hz/60Hz
6. Connector for lamp
7. Power cable
The supply unit is intended for power supplies above 220V.
To run it on lower voltages a step-down transformer is required
(e.g. 230V to 110V).
Observe the safety regulations on page 5.
Connect the cable from the lamp housing to the supply unit
Engage “L1” or “L2” in accordance with the inscription
on the burner socket (see page 12)
Engage50Hzor60Hz
Connect the power cable to the supply unit and to the grid
Make a note of the running-time meter reading.
Supply unit for lamp housing 106Z
with 100W mercury-vapour burner
1. On/off switch
2. Pilot lamp LAMP
3. Pilot lamp TEMP
4. Running-time meter
5. Pilot lamp SAFETY
Observe the safety regulations on page 5
Connect the cable from the lamp housing to the supply unit
Secure the lamp cable.
Connect the power cable to the supply unit and to the grid
Make a note of the running-time meter reading.
2 1 6 7
543
Input type
220V/240V±10%
220V~50/60Hz
Hg50WL1/L2
Osram no. 0-958
Max. power 320VA
Type ebq 100dc
Input voltage 90V to 250V AC, max 265VA
Inputfrequency 48Hzto63Hz
Protection class IP20
Series no. G 34240090
Fuses
The fuse holder is secured in the holder with spring clips
on both sides.
Insert a screwdriver behind the spring clips and extract
the fuse holder.
Insert the fuses (2×T3,15AH250VF1/F2).
Refit the fuse holder
1
3
2
4 5