Invacare XPO100 Oxygen Equipment User Manual


 
SECTION 6—OPERATING INSTRUCTIONS
Part No 1148112 23 Invacare XPO
2
Charging the Supplemental Battery Only
NOTE:Forthisprocedure,refertoFIGURE6.4onpage 24.
1. ConnecttheAC(Detail“A ) orDC(Detail“B”)power
adaptertothesupplementalbatterypack.
2. ConnecttheotherendoftheAC(Detail“A ) orDC(Detail
“B”)poweradapterintotheappropriatepowersource.
NOTE:Ifcompletelydischarged,thesupplementalbatterywillfully
chargeinapproximately3to4hours.SupplementalBatteryGauge
indicatorlightsDONOTilluminateduringcharging.
Supplemental Battery Status Button
NOTE:Forthisprocedure,refertoDetail“C”inFIGURE6.4on
page 24.
NOTE:Pressingthebatterystatusbuttonwhilechargingturnsthe
batterygaugeOnandshowsthestatusofthebatterycapacity.The
batterygaugealsoflashes“thelast”orhighestlightbartoshowthat
thebatterypackischarging.Theflashingbarshowsthebatteryis
actuallybeingcharged.Thelitbarsshowthestatusofthebattery
capacitybeingrestored.