Invacare XPO100 Oxygen Equipment User Manual


 
SECTION 5—TYPICAL PRODUCT PARAMETERS
Part No 1148112 15 Invacare XPO
2
Oxygen Concentration:*
*Based on an atmospheric
pressure of 14.7 psi (101 kPa) at
70°F (21°C)
87% to 95.6%, after initial warm-up
period (approximately 5 minutes)
Oxygen Flow: Pulse flow delivery. Bolus Volume Ranges
From 300 - 840cc settings 1 to 5.
Dimensions: 10” high x 7” wide x 4” deep (25.4 cm
high x 17.8 cm wide x 10.2 cm wide)
Weight: 6.0 lbs (2.72 kg); 1.3 lbs (0.68 kg) for
supplemental battery
Battery Duration (times are
approximate):
Setting 1 = 3.5 hrs
Setting 2 = 2.5 hrs
Setting 3 = 2.0 hrs
Setting 4 = 1.5 hrs
Setting 5 = 1.0 hrs
Battery Recharge Time:
NOTE:Rechargetimeincreasesif
batteryischargingwhileunitis
running.
4 hours
Humidity Operating Humidity: 15% to 60% non
condensing
Storage Humidity: up to 95% non
condensing
Temperature Range:
(All power sources)
Operating temperature: 41°F to 95°F
(5°C to 35°C)
Storage temperature: -2°F to 140°F
(-20°C to 60°C)
Extended Temperature Range:
(Using AC or DC adapters)
Operating Temperature:
95°F to 104° (35°C to 40°C)
Continuous use - all settings
Extended Temperature Range
(Using Internal Battery)
Operating Temperature:
95°F to 104° (35°C to 40°C)
Continuous use - settings 1, 2 and 3
45 minutes (max) - setting 4
30 minutes (max) - setting 5