Invacare PH904A Mobility Aid User Manual


 
SECTION 7—TILT
Compass™SPT™ 54 Part No 1122134
7. When the seat reaches the desired angle, slow let go of the trigger release levers.
8. To lower the seat parallel with the floor, reverse the above sections.
FIGURE 7.1 Engaging Tilt-In-Space
Adjusting Trigger Release Cables
NOTE: For this procedure, refer to FIGURE 7.2 on page 56.
14 to 20-Inch Wide Wheelchairs
ƽ WARNING
DO NOT adjust the trigger release cables while the cables are coiled up. The adjust-
ment will be too tight when the cable is straightened and the gas cylinder will not
lock and hold its position, possibly causing injury.
1. To adjust the left gas cylinder, squeeze the trigger release lever of the right cable
assembly and try to tilt the wheelchair.
NOTE: The wheelchair should not tilt.
2. If the left side of the wheelchair tilts without squeezing the trigger release lever of the
left cable assembly, perform the following steps:
A. Loosen the cable locking nut that secures the cable adjusting nut in place.
B. Turn cable adjusting nut one of two ways:
NOTE: If the cable is too loose, the tilt mechanism will not work.
ƽ WARNING
DO NOT over tighten the cable adjusting nut on the cable. Over tightening the
cable adjusting nut on the cable WILL unlock the tilt mechanism, possibly causing
injury.
Back Cane
Trigger Release
Lever