CARE AND MAINTENANCE
FUNCTIONAL CHECK
CHECKING FOR LEAKS
1. Close the valve on the oxygen cylinder.
2. Depressurize the EVOLUTION
TM
MOTION by inhaling several times using the nasal cannula. The
gauge indicator should drop to zero.
3. Check that all screwed unions and tube connections are tight. If necessary, tighten them by hand.
*
NOTE: Do not use a tool.
4. Ensure that the unit is set to the “PULSE” position.
5. Slowly open the valve on the oxygen cylinder until the needle in the gauge indicator no
longer moves.
6. Close the cylinder valve again.
7. Observe the needle in the gauge indicator for approximately one minute.
• If the needle remains in its position, everything is OK.
• However, if the contents indicator shows a continuous decrease in pressure, there is a leak in
the system. In this event, contact your Home Care Provider.