Evolution Technologies OM-900M Oxygen Equipment User Manual


 
M2 36 liters 3.0 1.5 1.0 0.8 0.6 0.5 0.3
M4(A) 113 liters 9.4 4.7 3.1 2.4 1.9 1.6 0.9
M6(B) 164 liters 13.7 6.8 4.6 3.4 2.7 2.3 1.4
ML6 171 liters 14.3 7.1 4.8 3.6 2.9 2.4 1.4
M7 198 liters 16.5 8.3 5.5 4.1 3.3 2.8 1.7
M9(C) 246 liters 20.5 10.3 6.8 5.1 4.1 3.4 2.1
D 425 liters 35.4 17.7 1.8 8.9 7.1 5.9 3.5
E 680 liters 56.7 28.3 18.9 14.2 11.3 9.4 5.7
16
OXYGEN CYLINDER DURATION
Because the total delivery of oxygen via the EVOLUTION
TM
MOTION conserver is related to
breathing rates and activity level, it is user adaptive in that the total oxygen delivered per minute
will automatically adjust with user need. These cylinder durations can vary widely depending on
whether more time is spent in Rest or Active mode and depending on breath rate.
To use the duration chart below, you can use an average of your Rest setting and your Active
setting if you are using your device equally between the two setting options. For example, if your
Rest setting is “2” and your Active setting is “4”, then your tank duration on an M6(B) cylinder at
20 breaths a minute would be approximately 4.6 hours. If you spend more time in rest mode, your
duration would be higher, and if you breathe faster than 20 breaths per minute while active, the
duration may be lower. This chart is intended to be used as a guide only and is an estimation of
use times. Actual results may vary.
SETTING 1 2 3 4 5 6 Continuous
Flow
2 LPM
CYLINDER TYPE CYLINDER VOLUME
Estimated Cylinder Duration in Hours
(Based on 20 breaths per minute)
11.8