Evolution Technologies OM-900M Oxygen Equipment User Manual


 
15
ASSEMBLY AND USE
OPERATING INSTRUCTIONS (Cont.):
The Active setting should be used during exercise or activity and the Rest setting should be
used when inactive or at rest. The Rest and Active settings should be selected in
accordance with a physician’s prescription or with a physician’s approval.
The motion sensor will select Rest or Active mode in accordance with the below table.
ACTIVITY RESULT
Motion occurs for less than 2 seconds Remain in Rest mode
Motion occurs for more than 2 seconds Switch to Active mode
If motion occurs between 2 and 4 seconds Revert to Rest mode 10 seconds after
motion ceases
If motion occurs for more than 4 seconds Revert to Rest mode 50 seconds after
motion ceases
*
NOTE: To help prevent possible damage to the unit, keep the EVOLUTION
TM
MOTION conserver in a carrying bag. Several
bags are available for use with different cylinder sizes and configurations.
STEP 6: When finished using the system, turn off the oxygen supply cylinder
valve and continue breathing through the nasal cannula until no
further oxygen is detected.
STEP 7: Remove the nasal cannula.
STEP 8: When not in use, store in a clean, dry location.