Evolution Technologies OM-900M Oxygen Equipment User Manual


 
OPERATING INSTRUCTIONS:
STEP 1: Make sure the EVOLUTION
TM
MOTION conserver is set to the “PULSE” position. Battery
saving technology has eliminated the need for an “OFF” position. The unit will go to “Sleep
Mode” after three minutes of non-use. The unit will turn on again when an inspiration is
detected or the Selector Button is pressed.
STEP 2: To reduce the risk of rapid oxygen recompression and fire, OPEN THE CYLINDER VALVE
SLOWLY and completely so that the pressure gauge moves slowly as it indicates the
cylinder pressure.
STEP 3: Listen for leaks. If a leak is present, close the cylinder valve, check the seal washer, and
reinstall. If the leak persists, DO NOT USE THE EQUIPMENT. Contact your supplier for repair.
STEP 4: To select the Rest and Active settings on the EVOLUTION
TM
MOTION OM-900M conserver
(1-6) that are equivalent to your oxygen prescription, press and hold the REST Selector Button
on the Selector Knob until the appropriate LED display light is illuminated. Next, press and hold
the Active Selector Button on the Selector Knob until the appropriate LED display light is
illuminated.
*
NOTE: To check your current settings, press and release the appropriate Rest or Active Selector Button
*
NOTE: The Active setting can only be set equal to or greater than that of the Rest setting.
Because the EVOLUTION
TM
MOTION stores the settings from the last
time it was used, you only have to complete this step before your first
use or after you have made a change to your normal settings.
STEP 5: Place the nasal cannula into position with the prongs in the nostrils
and begin breathing. [Fig. G] The EVOLUTION
TM
MOTION conserver
will now start to deliver oxygen. The amount of oxygen delivered
is determined by the setting. Adequate oxygen delivery will be
achieved because of the precise time in the breathing cycle that the
pulse of oxygen is delivered.
14
ASSEMBLY AND USE
Fig. G
Proper Positioning of
Nasal Cannula