Emerson Process Management 3000 Oxygen Equipment User Manual


 
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
2-18 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
+
+
+
+
+
+
+
+
+
-
-
-
-
-
-
-
-
-
BK
WH
J9
J8
J3
J2 J1
TRIAC RELAY
STACK
TC
ANALOG
HEATER
PROBE
MV
PROBE
MV
PROBE
TC
PROBE
TC
AD590
(INTERNAL
WIRING)
WHBK
OR CELL+
YE HTR TC +
BL CELL -
RD HTR TC
-
BK
WH
GN
STACK
TC
PROBE
HEATER
RH
N
L
LINE
VOLTAGE
GN/YE
TOP
BOTTOM
4 TWISTED PAIR SHIELDED
#22 AWG BY CUSTOMER
2 TWISTED PAIR SHIELDED
#22 AWG BY CUSTOMER
(OPTIONAL)
HPS 3000 INTERFACE MODULE
LINE
1
2
2
7
11
1
4
1
2
3
4
5
6
7
8
9
10
WORLD CLASS PROBE
ALWAYS DISCONNECT LINE VOLTAGE
FROM HEATER POWER SUPPLY
BEFORE CHANGING JUMPERS.
JUMPER
CONFIGURATION
LINE VOLTAGE
SELECTION
JUMPER
(INSTALL)
JUMPER
(INSTALL)
HEATER
POWER
ON
120 V.A.C./100 V.A.C.
220/240 V.A.C.
JM4, JM1
JM5
REMOTE
ON
REMOVE JM2
INSTALL JM2
ELECTRONICS
SELECTION
JUMPER
PROBE HEATER
VOLTAGE SECTION
REMOVE JM3, JM6DIGITAL
(NEXT GENERATION)
WORLD CLASS PROBE JM7
RELAY WIRE IS OPTIONAL, RELAY CAN BE BYPASSED
WITH JUMPER JM-2 IF NOT WIRED TO THE IFT.
STACK TC WIRING AS REQUIRED.
ALL WIRES #16-#22 AWG TWISTED PAIR WITH SHIELD
BY CUSTOMER EXCEPT AS NOTED.
STANDARD PROBE CABLE BETWEEN PROBE AND
HPS BY ROSEMOUNT.
REMOVE JM1 ON INTERCONNECT BOARD.
REMOVE JM6 ON MICROPROCESSOR BOARD.
IF RELAY WIRE OF NOTE 1 INSTALLED THEN
REMOVE JM2 ON HPS 3000.
IF STACK TEMPERATURE NOT USED.
IF MPS 3000 NOT USED.
1 RELAY PER PROBE AVAILABLE FOR CALIBRATION
STATUS INDICATION (48 V max, 100 mAmax).
100 V.A.C. REQUIRES TRANSFORMER PART
NUMBER 1M02961G02.
JUMPER JM7 INFORMATION APPEARS IN FIGURE 2-9.
NOTES
12345678
BL
PROBE MV -
PROBE MV +
PROBE TC +
PROBE TC -
OR
YE
RD
GN
E
R
H
WH
BK
GN CELL -VE
OR CELL +VE
YE CHROMEL
RD ALUMEL
GN
BK
BK
}
HEATER
PROBE JUNCTION
BOX WIRING
A
B
11
22540001
Figure 2-14. Wiring Layout for Complete IFT 3000 System with HPS (Sheet 1of 2)