Emerson Process Management 3000 Oxygen Equipment User Manual


 
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
2-14 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
OUTPUT JUMPER
HPS Remove JM6
Probe (No HPS) Install JM6
ANALOG OUTPUT
(Condition during
microcontroller failure) Jumper
Output = zero Install JM7
Output = maximum Remove JM7
(See Figure 2-10 for jumper locations.)
Figure 2-9. Microprocessor Board
Jumper Configuration
c. Analog Output and Relay Output
Connections
1. The microprocessor board has a se-
lector for voltage or current operations.
Figure 2-10 shows switch orientation.
In voltage mode, output is 0-10 V. In
the current mode, the output can be
configured from the SETUP menu to
be 0-20 mA or 4-20 mA.
2. The analog output and relay outputs
are programmed by the user as
needed. The analog output is typically
sent to recording equipment such as
chart recorders. Relay outputs are typi-
cally sent to annunciators.
3. Relays K1 and K2 are user configur-
able from the probe SETUP sub-menu
(Table 5-5). Typically these are used to
indicate O
2
values above or below
specified tolerances. OK relay is ener-
gized when unit is functioning properly.
4. All wiring must conform to local and
national codes.
5. Connect the analog output and relay
outputs as shown in Figure 2-12.
2-3 HEATER POWER SUPPLY INSTALLATION
a. Mechanical Installation
The outline drawing of the heater power
supply enclosure in Figure 2-13 shows
mounting centers and clearances. The
NEMA 4X enclosure is designed to be
mounted on a wall or bulkhead. The heater
power supply should be installed no further
than 150 feet (45 m) from the probe. The
heater power supply must be located in a
location free from significant ambient tem-
perature changes and electrical noise. Am-
bient temperature must be between -20°
and 140°F (-30° and 60°C).