11
MODEL 1056DP SECTION 4.0
PROGRAMMING PROFIBUS
SECTION 4.0
PROGRAMMING PROFIBUS
4.1 CHANGING PROFIBUS ADDRESS AT THE DEVICE
Properly wired and configured Profibus Model 1056 units can be accessed by the Profibus master after a GSD file
is downloaded to the Master. See Appendix A for details on the GSD file available from Rosemount Analytical.
The only programmable setting that may need to be changed is the default Profibus address or Profibus address
that was entered upon initial Quick Start routine.
Note that the Profibus device network address can also be changed from the master unit.
4.1 CHANGING PROFIBUS ADDRESS AT THE DEVICE
Profibus-configured Model 1056 units (Model number 1056-0X-2X-3X-DP) include a pre-installed Profibus com-
munications board that is recognized by the main PCB.
To change the Profibus address,
1. Press MENU. Then select Program. The following
screen will appear.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Program
Profibus
Outputs
Alarms
Measurement
2. Select Profibus. The following screen will appear.
3. Using the keypad, enter a numeric Profibus network address from 000-125 for the Model 1056 Profibus
unit already commissioned.
S1: 1.234µS/cm 25.0ºC
S2: 12.34pH 25.0ºC
Profibus
Address: 126