6
2. Secure the purple Profibus cable with a cable
gland fittings and nut from the outside of the enclo-
sure to ensure a proper seal. Note that the cable
gland fittings (PN 23554-00 - Cable Gland Kit,
Quantity 5) does not require the securing nut inside
the enclosure to properly install the fitting. Simply
thread the fitting into the grounding plate inside the
enclosure and tighten.
Once the gland fitting is tightly screwed into the
enclosure and internal grounding plate, only the
external nut needs to be tightened to properly seal
the Profibus cable.
MODEL 1056DP SECTION 3.0
INSTALLATION AND WIRING
3.2 Wiring Directly To A Profibus Network
To connect a Profibus-configured Model 1056 to a Profibus network, the conventional method is to wire a stan-
dard purple Profibus cable directly to the Profibus board inside the instrument. The Profibus board is designed
to accommodate direct wiring of a Profibus cable to a removable 4-lead terminal block on the Profibus board.
1. Feed the purple Profibus cable through a cable gland fitting. Install the cable gland fitting into the enclosure
opening on the leftmost side of the enclosure nearest the front of the door hinge. See photo. Note that the
power cable should be installed in the leftmost side of the enclosure farthest from the front of the door hinge
– as shown in the photo below.
Proper enclosure opening locations for
cable installation
Cable gland fittings secured on
internal grounding plate.