Curtis 40 Wheelchair User Manual


 
51
Curtis enAble
®
40-Series Manual, Rev. B
ATTENDANT CONTROL DIAGNOSTICS
The LEDs on the attendant control are used to communicate information
about faults, warnings, and battery charging.
The Mode LED, the BDI LED, and the LEDs in the chair icon signal
warnings by flashing. For more information, turn on the user handcontrol and
refer to Table 1.
8 — DIAGNOSTICS & TROUBLESHOOTING
Table 2: ATTENDANT CONTROL FAULT & WARNING INDICATORS
SIGNAL FAULT/WARNING/INFO COMMENTS
MODE LED
(RED)
(Various faults and warnings.) See display on the user
handcontrol to ascertain
which fault is present; then
see Table 1 for further
information.
CHAIR ICON LEDs
(GREEN)
Flashing Seat Recline Fault. See Table 1.
Seatback LED
Flashing
Seat Lift Fault. See Table 1.
Seat LED
Flashing
Left or Right Leg Fault. See Table 1.
Leg LED
All the chair
Actuator Over/Under Temperature See Table 1.
LEDs flashing Warning.
BDI LED
Flashing Red Undervoltage. See Table 1.
Flashing Green Overvoltage. See Table 1.
Flashing Red, Charging battery; inhibit.
Amber, Green
in Sequence