Braun NUVL855RM24 Wheelchair User Manual


 
Page 33
Manual Operating Instructions
Release Valve
maximum
30 inch lbs
minimum
15 inch lbs
seats
(stops)
O
P
E
N
C
L
O
S
E
approximate
1/16” intervals
Valve
Tightening
Specication:
Once valve
seats (stops),
tighten 15 to 30
inch pounds as
shown.
Note: Close backup pump
release valve securely be-
fore operating electric pump.
A
Note: The lift platform must be
pushed back into its carriage
compartment all the way before
reverting back to normal (pow-
ered) operation. When the lift is
fully extended manually, it does
not activate the proper switches
for normal operation. Returning
(moving) the lift in fully allows for
proper switch activation.
If you experience power or
equipment failure, refer to the
Manual Operating Instructions
to operate the lift. Instructions
and illustrations are provided
for all steps that differ from
standard lift operation proce-
dures. Manual Instructions
Decal 31409 (posted in pump
module box) provides manual
operating instructions also.
Release Valve
The manual hand pump is used
to lower and raise the platform
and outer barrier.
Manual Hand Pump