What You See What It Means What You Should Do
or higher
Meter has detected an
error; accurate result is
not possible.
Turn the meter off, then on
using , then retest. If the
error remains, contact Bayer
Customer Service.
The display
goes blank when
the meter is
turned on.
1. You may have
accidentally pressed
the
button.
2. Meter has turned
off after 3 minutes
without use or after
10 minutes of no use
while waiting for you
to apply blood or
control solution.
1. Press the
button again.
2. This is a normal expected action
that saves battery life. Press
to turn on again.
Meter will not
turn on.
1. Battery life has ended.
2. Battery holder is
loose or battery is
installed with the
+ side down.
1. Replace the battery (see page 35).
2. Check the battery and battery
holder to be sure they are
tightly in place and correctly
installed (see page 37). If the
display is still blank, replace the
battery (CR2032).
Meter fails to
start countdown
after blood is
applied.
1. Not enough blood in
test strip.
2. There may be debris
inside the meter.
3. Meter failure.
1. Discard the used test strip. Retest
using a round drop of blood.
2. Inspect the inside of the meter
(see page 34) and remove
debris if present. Retest.
3. Contact Bayer Customer Service.