Bayer HealthCare Breeze 2 Blood Glucose Meter User Manual


 
What You See What It Means What You Should Do
Low Battery symbol
flashes if battery power
is low.
Replace the battery (CR2032)
within 20 readings or one week.
Battery life has ended. Replace the battery (CR2032).
Temperature is outside
the operating range of
the meter (too hot or
too cold).
Allow the meter to reach room
temperature.
Meter does not
recognize the test disc.
Remove the test disc and inspect
the meter for foreign material.
Rotate the disc and reinstall it. If
error repeats, try inserting a new
disc. If error continues, contact
Bayer Customer Service.
Test strip was removed
during the test, or the
release button was
pressed during the test.
Take care to not press the release
button or remove the test strip
during the test. Release the used
test strip and repeat the test (see
page 10). If the error continues,
contact Bayer Customer Service.
(with continuous
beep)
The meter was open
when the meter handle
was pulled.
Close the meter and push the
handle in.
41