Agilent Technologies 6000 Series Respiratory Product User Manual


 
Performance characteristics (continued)
General characteristics (continued)
Trigger out When Triggers is selected (delay ~17 ns)
0 to 5 V into high impedance
0 to 2.5 V into 50 Ω
When Source Frequency or Source Frequency/8* is selected
0 to 580 mV into high impedance
0 to 290 mV into 50 Ω
Max frequency output: 350 MHz (in source frequency mode when terminated in 50 Ω)
125 MHz (in source frequency/8 mode when terminated in 50 Ω)
10 MHz ref in/out TTL out, 180 mV to 1 V amplitude with 0 to 2 V offset
Kensington lock Connection on rear panel for security
Power requirements
Line voltage range 100-120 V, 50/60/400 Hz; 100-240V, 50/60 Hz auto ranging
Line frequency 50/60 Hz, 100-240 VAC; 400 Hz, 100-120 VAC
Power usage 120 W max
Battery option – BAT 2+ hours between charges, battery-low indicator at 20%
Battery capacity after repeated charging: 80% after 300 cycles
Non-operating temperature: –20 °C to 60 °C
Operating temperature: 0 °C to 50 °C
Power consumption is 67-75 Watts with optional N5429A DC Power adapter
Environmental characteristics
Ambient temperature Operating 0 °C to +55 °C; non-operating –40 °C to +70 °C
Humidity Operating 95% RH at 40 °C for 24 hr; non-operating 90% RH at 65 °C for 24 hr
Altitude Operating to 4,570 m (15,000 ft); non-operating to 15,244 m (50,000 ft)
Vibration Agilent class B1 and MIL-PRF-28800F; class 3 random
Shock Agilent class B1 and MIL-PRF-28800F; class 3 random; (operating 30g, 1/2 sine,
11 ms duration, 3 shocks/axis along major axis, total of 18 shocks)
Pollution degree Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation must be expected.
Indoor use Rated for indoor use only
Other
Measurement categories CAT I
Regulatory information Safety IEC 61010-1:2001 / EN 61010-1:2001
Canada: CSA C22.2 No. 1010.1:1992
UL 61010B-1:2003
Supplementary information The product herewith complies with the requirements of the Low Voltage Directive
73/23/EEC and the EMC Directive 89/336/EEC, and carries the CE-marking accordingly.
The product was tested in a typical configuration with HP/Agilent test systems.
* Source Frequency/8 is supported on 300 MHz to 1 GHz 6000 Series only.
24