A&D SW-150KL Scale User Manual


 
37
Command examples (“_” indicates “Space” (20H).)
The examples below are for the function setting “aCk 1 “ (Reply to commands).
The address ## = 23.
Request a weight data.
Command
@ 2 3 Q C
R
L
F
Reply
@ 23 S T , +0012. 345_kg C
R
L
F
Stable data
@ 23 U S , +0007.890_kg C
R
L
F
Unstable data
@ 23 O L , +9999.999_kg C
R
L
F
“E” display
Zero the scale. (No reply for the function setting “aCk 0 “.)
Command
@ 2 3 Z C
R
L
F
Reply
@ 2 3 Z C
R
L
F
The scale is in a condition that zero operation is possible.
Send the current LO limit value.
Command
@ 2 3 ? L 1 C
R
L
F
Reply
@ 2 3 L 1 , +000200C
R
L
F