22
12. FUNCTIONS
The scale has function settings to specify the scale performance.
The parameters set in the function settings are maintained even if the power is turned
OFF.
12.1. Setting the parameters
1. Turn the power OFF.
2. Press and hold the TARE switch and press the
ON/OFF switch to turn the power ON. The software
version is displayed.
“∗∗∗” indicates the software version number.
3. After about one second, the item is displayed.
4. Change the item or parameter using the following switches.
UNITS To display the next item.
PRINT To increase the value of the blinking
digit by one (to change the parameter).
When the parameter is changed, the
STABLE indicator turns off.
5. Press the TARE switch to store the setting value.
After “end”, the power is turned OFF automatically.
To cancel the setting procedure without storing the value, press
the ZERO switch. The power is turned OFF automatically.
12.2. To restore the function settings to the
factory setting values
1. Turn the power OFF.
2. Press and hold the TARE switch and press the
ON/OFF switch to turn the power ON. Only release the
ON/OFF switch and hold the TARE switch until
“Clrf no” with “no” blinking is displayed.
3. Press the PRINT switch.
“Clrf no” changes to “Clrf go” with “go” blinking.
4. When “Clrf go” is displayed, press the TARE switch.
The factory setting values are restored. After “end”, the power
is turned OFF automatically.
To cancel the restoring procedure, press the ZERO switch.
The power is turned OFF automatically..
Item
Parameter
While
holding down, press
Press
Press
When changed
Press
Press