Operating Instructions ….
Operating the PM-D3 Washer Disinfector
1. Apply power to the machine and press the DOOR RELEASE button to open the door.
2. Load the instruments to be cleaned into the Large Basket, Small Instrument Basket,
Plate Rack, and Instrument Trays as required. Slide Large Basket into chamber.
Do not overload the Large Basket so that water spray is obstructed.
!
Water is sprayed up from the bottom of the chamber so large items, such as
trays and bowls, should NOT be placed flat on the bottom of the Large Basket
as this will obstruct the water spray.
Note:
Whenever hollow instruments are included in the load it is important that any spare
luer connectors are blanked off with the caps supplied with the connectors. This will
ensure that an effective irrigation flow is achieved through each hollow device and not
by-passed through the spare luer connectors.
3. Take a sachet of Decono~Zyme Gel, tear or cut off one corner of the pack and empty
the entire contents into the door cavity marked ‘DETERGENT’ and close door.
4. Observe the messages shown on the LCD DISPLAY and follow this procedure:
If this message is displayed then
refer to the “AWT Fluid
Replenishment” section on page
12 of this manual.
Load Descaler
Press START
Press START button once
If AWT Fluid has already been
added then continue. If not, then
refer to the “AWT Fluid
Replenishment” section on page
12 of this manual (The descaler
messages will only be shown at
very first power-up or when AWT
Fluid is low.
)
IF Descaler IN
Press START
Press START button once
If Decono~Zyme detergent has
already been loaded then
continue. If not, then return to
step 3.
Load Instrumen s t
Load Detergent
Press START button once
Issue 1 06/10/2004 7