A SERVICE OF

logo

Prestige PM-D3 Washer Disinfector Section 2
Technical Specification ….
General:
Model: PM-D3 Washer Disinfector
Description: Washer Disinfector, Class IIa
Type: Cabinet (single chamber) machine (type 1)
Purpose: To clean or clean/disinfect surgical instruments, hollow instruments,
and associated products
Operating Process: Moist heat thermal disinfection
Applied Standards: BS2745: part 1: 1993
BS2745: part 3: 1993
EN60601-1-2: 1993
EN61010-2-045 (2001)
EN61326: 1997 & A1: 1998
Wash Cycle Parameters:
Wash temperature: 42°C. to 45°C.
Disinfection Conditions: Moist heat
@ 80°C. to 85° C.
Disinfection Period Band: 80 - 85° C. for a minimum period of 1 minute
Holding Time (prog 1) 2 minutes
@ Disinfection Conditions
Holding Time (prog 2) 10 minutes
@ Disinfection Conditions
Equilibration Time: approximately 3 minutes
Plateau Period: approximately 5 minutes
Min. Temp. During Cycle: determined by temperature of water supply
Max. Temp. During Cycle: 85°C.
Accuracy of Temp. Measurement: < 2% @ 85°C.
Vol. of Dispensed Chemicals: AWT Fluid: 2ml (approx)
Decono~Zyme: 30ml maximum
Required Services:
Electricity Supply:
Supply Voltage (230v): single phase 230v (nominal) AC
Supply Voltage (110v): single phase 110v (nominal) AC
Supply Frequency: 50/60 Hz
Lead Length: 1.83M (6 feet)
Water Supply:
Supply Connection: ¾” BSP Male Union
Acceptable Pressure Range: 1 – 3 BAR (15 – 45psi)
Effect of Lower Pressure: increased filling time leading to extended total cycle
time
Acceptable Flow Range: not less than 2 litres/min
Effect of Reduced Flow: increased filling time leading to extended total cycle
time
Volume Used per Cycle: 9 litres maximum
Acceptable Temp. Range: 15 - 25°C.
Hardness: not more than 5mg/L (5ppm) CaCo
3
Issue 1 06/10/2004 16