AWT Fluid Replenishment ….
When full, the AWT Fluid reservoir contains sufficient anti-scaling fluid for around 40 wash
and disinfect cycles. On completion of 25 cycles, the LCD DISPLAY will indicate that more
fluid is required and the unit will not proceed with a new wash and disinfect cycle until you
confirm that it has been added.
To add AWT Fluid, the following procedure applies:
1. Apply Power to the Machine. Press DOOR RELEASE to open door.
2. Unscrew the reservoir cap which is located on the inside rear of the wash
chamber.
3. Using the AWT dispenser bottle, slowly top up the reservoir until it begins to
overflow.
4. Refit the cap on the reservoir.