A SERVICE OF

logo

Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Programming Guide 2-35
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Keyword/Syntax Parameter Form Std/
New
Comments
[:SENSe] (Cont.)
:ROSCillator (Cont.)
:SOURce
:AUTO
:TINTerval
:ARM (HP 53131A
and HP 53132A with
S/N prefix below
3646)
[:STARt]
:SLOPe
:SOURce
:STOP
:SOURce
:TIMer
:TOTalize
:ARM
[:STARt]
:SLOPe
:SOURce
INTernal | EXTernal
<Boolean>
See Table 2-2A on
page 2-38 for HP
53132A (with S/N
3646 and above)
TINTerval
ARM commands)
POSitive | NEGative
IMMediate | EXTernal
IMMediate | TIMer
<numeric_value> [S]
POSitive | NEGative
EXTernal | IMMediate
Std
Std
New
New
New
New
New
New
New
New
New
New
New
New
New
Sets the selection of a reference timebase.
Sets the enable for automatically selecting a reference
timebase.
Subtree. Controls the time interval (including Time
Interval, Rise Time, Fall Time, Dutycycle, and Pulse
Width functions) measuring capabilities of the
instrument.
Subtree. Synchronizes the time interval start and stop
arm with events.
Subtree.
Sets the slope of the external start arm signal used in
external arming time interval measurements (including
Time Interval, Rise Time, Fall Time, Dutycycle, and
Pulse Width functions). Only applies when
[:SENS]:TINT:ARM[:STAR]:SOUR EXT is selected.
Sets the start arm for time interval (including Time
Interval, Rise Time, Fall Time, Dutycycle, and Pulse
Width) measurements.
Subtree.
Sets the stop arm (i.e., the delay) for Time Interval
measurements.
Sets the time used to delay the stop arm for Time
Interval measurements. Only applies when
[:SENS]:TINT:ARM:STOP:SOUR TIM is selected.
Subtree. Controls the totalize measuring capabilities of
the instrument.
Subtree. Synchronizes totalize start and stop arm with
events.
Subtree.
Sets the slope of the external start arm signal used in
external arming totalize measurements. Only applies
when [:SENS]:TOT:ARM[:STAR]:SOUR EXT is selected.
Sets the start arm for totalize measurements.