Chapter 4 Command Reference
Introduction
Programming Guide 4-21
Stats
:CALCulate3:AVERage:COUNt:CURRent?
Queries the current count (that is, the number of data values collected for
statistical computation).
• Numeric data transferred as ASCII bytes in <NR1> format.
• Range is 0 to 1,000,000.
• If :CALC3:AVER[:STATe] is OFF, error -221 is generated.
• If the current measurement is Totalize or Voltage Peaks, 0 is returned and
error -221 is generated.
• Query only.
• No statistics results exist until the :CALC3:AVER:COUN:CURR? is equal to
the specified :CALC3:AVER:COUN.
Stats
:CALCulate3:AVERage[:STATe] <Boolean>
Sets or queries the statistics post-processing enable.
• Single ASCII-encoded byte, 0 or 1.
• A value of 0 indicates OFF; a value of 1 indicates ON.
• *RST: OFF
• When this enable is ON, and :TRIG:COUN:AUTO is ON, and
[:SENS]:FUNC[:ON] is not Totalize or Voltage Peak function,
then :INIT[:IMM] initiates a complete block of measurements.
See :TRIG:COUN:AUTO in this chapter for specifics.
• When this enable is OFF, :INIT[:IMM] always initiates a single measurement.