Table 3. Flow Rate and Flow velocity
(SI unit)
Unit: m
3
/h
Size
(mm)
Flow rate
0.1 m/s 0.3 m/s 1.0 m/s
3.0
m/s
12
m/s
15
0.06362 0.1908 0.6361 1.908 7.634
25
0.1767 0.5301 1.767 5.301 21.21
32
0.2895 0.8686 2.895 8.686 34.74
40
0.4523 1.357 4.523 13.57 54.29
50
0.7067 2.120 7.067 21.20 84.82
65
1.195 3.583 11.95 35.83 143.4
80
1.809 5.428 18.09 54.28 217.1
100
2.827 8.482 28.27 84.82 339.3
125
4.417 13.25 44.17 132.5 530.1
150
6.361 19.08 63.61 190.8 763.4
200
11.31 33.93 113.1 229.3 1,357
250
17.67 53.01 176.7 530.1 2,121
300
25.45 76.34 254.5 763.4 3,054
350
34.64 103.9 346.4 1,039 4,156
400
45.23 135.7 452.3 1,357 5,429
450
57.25 171.7 572.5 1,717 6,871
About establishment environment
Do not store or install the flowmeter:
• Where there is direct sunlight.
• Where excessive vibration or mechanical shock
occurs.
• Where high temperature or high humidity
conditions exist.
• Where corrosive atmospheres exist.
• Places that can be submerged under water.
• When placing the flowmeter temporarily on the
floor, support it, i.e., with a block so that the
flowmeter does not topple over.
In areas like the following, there may be the case that
infrared switches do not function correctly. (If these
are unavoidable, use an appropriate cover.)
(1) Where unit (operation panel) is exposed to direct
sunlight, reflection of light onto window pane
and diffused light reflection.
(2) Where smoke and steam may occur.
(3) Where exposed to direct snow, ice or mud.
Ordering Information
1. When ordering the LF654 series flowmeters, refer
to Tables 4 and 5 (Type Specification Codes).
An entry must be made for each of the columns in
each of these tables.
2. Fluid characteristics:
(1) Type of fluid to be measured and its characteristics
(2) Fluid temperature
(3) Fluid pressure
(4) Electrical conductivity of the fluid
3. Measuring range
4. I/O function setting
5. Ordering scope
Flow calibration data
6. Other items
Specifications other than standard items
Consult a Toshiba representative when choosing
materials for lining, electrodes, and grounding rings.
Toshiba International Corp. (Houston) stocks
Hazardous location certification type
flowmeters as standard inventory.