Teledyne 3000PA Oxygen Equipment User Manual


 
3 Installation Model 3000PA
3-8
Teledyne Analytical Instruments
Note: The Remote Probe connector (paragraph 3.3.3) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously. If you have the C Internal valve
optionwhich includes additional zero and span gas inputs
the 3000P automatically regulates the zero, span and sample
gas flow.
Range ID Relays: Four dedicated Range ID relay contacts. The first
three ranges are assigned to relays in ascending orderLow range is as-
signed to Range 1 ID, Medium range is assigned to Range 2 ID, and High
range is assigned to Range 3 ID. The fourth range is reserved for the Air Cal
Range (25%).
Network I/O: A serial digital input/output for local network protocol.
At this printing, this port is not yet functional.
RS-232 Port: The digital signal output is a standard RS-232 serial
communications port used to connect the analyzer to a computer, terminal, or
other digital device. It requires a standard 9-pin D connector.
The data is status information, in digital form, updated every two
seconds. Status is reported in the following order:
The concentration in percent
The range in use (HI, MED, LO)
The span of the range (0-10 %, etc)
Which alarmsif anyare disabled (ALx DISABLED)
Which alarmsif anyare tripped (ALx ON).
Each status output is followed by a carriage return and line feed.
Four input functions using RS-232 have been implemented to date.
They are described in Table 3-1.
Table 3-1: Commands via RS-232 Input
Command Description
as<enter> Immediately starts an autospan.
az<enter> Immediately starts an autozero.
co<enter> Reports "Raw Cell Output" (current output of the sensor
itself) in µA. For example
Raw Cell Output: 99 µA.