SP-505 Rev. B
DEVILBISS 5-LITER SERIES DEVILBISS 3-LITER SERIES DEVILBISS LT SERIES
Catalog Number 505DZ/505DS 505KS/505KZ/505NS 303DZ/303DS/303DA 303IZ/303IS 501DZ 501IZ
Delivery Rate 1/2 to 5 LPM 1/2 to 5 LPM 1/2 to 3 LPM 1/2 to 3 LPM 1/2 to 5 LPM 1/2 to 5 LPM
Oxygen Percentage 1 LPM = 94% ± 3% 220-230V~, 50Hz 1/2-3 LPM = 93% ± 3% 220-240V~, 50Hz 1-4 LPM = 93% ± 3% 220-240V~, 50Hz
2-4 LPM = 95% ± 3% 1 LPM = 93% ± 3% 1/2-3 LPM = 93% ± 3% 5 LPM = 88% +3% -5% 1-3 LPM = 93% ± 3%
5 LPM = 93% ± 3% 2-4 LPM = 95% ± 3%
220-230V~, 60Hz 4 LPM = 90% ± 3%
5 LPM = 85% +6% -3% 1/2-3 LPM = 93% ± 3% 5 LPM = 80% +3% -5%
230V~, 60Hz 230V~, 60Hz
1 LPM = 92% ± 3% 1-3 LPM = 93% ± 3%
2-4 LPM = 95% ± 3% 4 LPM = 90% +5% -3 %
5 LPM = 92% ± 3% 5 LPM = 82% ± 3%
Maximum Recommended Flow 5 LPM 4 LPM 3 LPM 3 LPM 5 LPM 5 LPM
Sound Level 49 dBAAvg. 45 dBAAvg. 51 dBAAvg. 51 dBAAvg. 51 dBAAvg. 49 dBAAvg.
Outlet Pressure 8.5 psig (58.6 kPa) 8.5 psig (58.6 kPa) 8.5 psig (58.6 kPa) 8.5 psig (58.6 kPa) 8.5 psig (58.6 kPa) 8.5 psig (58.6 kPa)
Electrical Rating 115V~, 60Hz, 4.4Amp 220-230V~, 50Hz, 1.7 Amp 115V~, 60Hz, 2.1 Amp 220-240V~, 50Hz, 1.1 Amp 115V~, 60Hz, 5 Amp 220-230V~, 50Hz, 1.7 Amp
230V~, 60 Hz, 2.3 Amp* 220-230V~, 60 Hz, 1.2 Amp 230V~, 60 Hz, 1.7 Amp
Operating Voltage Range 95-132V~, 60Hz 187-253V~, 50Hz 95-132V~, 60Hz 187-264V~, 50Hz 95-132V~, 60Hz 187-264V~, 50Hz
195.5-253V~, 60 Hz* 187-253V~, 60 Hz 193-253V~, 60 Hz
Power Consumption 400 Watts Average 330 Watts Avg @230V, 50Hz 225 Watts Average 240 Watts Avg @230V, 50Hz 415 Watts Average 315 Watts Avg @230V, 50Hz
340 Watts Avg @230V, 60Hz 270 Watts Avg @230V, 60Hz 360 Watts Avg @230V, 60Hz
Weight 51 lbs. (23 kg.) 53 lbs. (24 kg.) 37 lbs. (17 kg.) 41 lbs. (19 kg.) 47 lbs. (21 kg.) 47 lbs. (21 kg.)
Dimensions 27.75”H x 16”W x 14”D 27.75”H x 16”W x 14”D 23”H x 13.87”W x 12”D 23”H x 13.87”W x 12”D 27.75”H x 16”W x 14”D 27.75”H x 16”W x 14”D
(70.5 x 40.6 x 35.6 cm) (70.5 x 40.6 x 35.6 cm) (58.4 x 35.2 x 30.5 cm) (58.4 x 35.2 x 30.5 cm) (70.5 x 40.6 x 35.6 cm) (70.5 x 40.6 x 35.6 cm)
Pressure Relief Valve 35 psig ± 3 psig 35 psig ± 3 psig 25 psig ± 3 psig 25 psig ± 3 psig 35 psig ± 3 psig 35 psig ± 3 psig
(241 kPa ± 20.7 kPa) (241 kPa ± 20.7 kPa) (172 kPa ± 20.7 kPa) (172 kPa ± 20.7 kPa) (241 kPa ± 20.7 kPa) (241 kPa ± 20.7 kPa)
Operating System Pressure Sensing/ Pressure Sensing/ Pressure Sensing/ Pressure Sensing/ Timed/ Timed/
Pressure Swing Pressure Swing Pressure Vacuum Pressure Vacuum Pressure Swing Pressure Swing
Operating Altitude Up to 1500 meters (4,921 feet) above sea level without degradation of performance across the operating voltage range.
(tested at 21ºC only) From 1500 meters (4,921 feet) to 4000 meters (13,123 feet), the end user should check with their Sunrise Medical provider for the suitability.
Operating Environment 5°C to 40°C, humidity range of 15% to 95%, non-condensing*
Recommended Operating 5°C to 40°C, humidity range of 15% to 95%, non-condensing* 5°C to 35°C, humidity range of 15% to 95%, non-condensing
Ambient Range (tested at sea level)
For units equipped with
an OSD, the visible “low oxygen”
85%±2% 82%+1%-3% 85%±2% 85%±2% Not Applicable Not Applicable
indicator will activate at the
following levels
Storage Conditions -40°C to 70°C, humidity range of 10% to 100%, including condensation
Equipment Class and Type Class II Equipment Double Insulated Type B Applied Part
Approval Body and CSA TUV CSA TUV CSA CSA
Safety Standard CAN/CSA-C22.2 IEC 601-1 + A1 + A2 CAN/CSA-C22.2 IEC 601-1 CAN/CSA-C22.2 CAN/CSA-C22.2
No. 601.1-M90 EN 60601 + A1 + A2 No. 601.1-M90 ISO 8359:1991 No. 601.1-M90 No. 601.1-M90
IEC 601-1 ISO8359: 1996 IEC 601-1 Complies with ISO 8359:1996 IEC 601-1 IEC 601-1
(505KZ excluded) (303DA excluded) (303IZ excluded)
EMC Compliance IEC801-2 through 5 IEC 601-1-2 IEC801-2 through 5 IEC 601-1-2 IEC801-2 through 5 IEC 601-1-2
To: CISPR 11/EN55011 CISPR 11/EN55011 CISPR 11/EN55011
Specifications subject to change without notice.
*505KS/505KZ/505NS = 25ºC maximum ambient for 60Hz rating.