Sunrise Medical 501 Oxygen Equipment User Manual


 
19
LANGUAGE
SP-505 Rev. B
ENGLISH
Symptom
A. Unit does not operate. Power light is off when the power
switch is On. Audible alarm is pulsing and Service Required
light is flashing.
B. (DeVilbiss 3 & 5 Liter Series) Unit does not operate but the Power
light is on when the power switch is On. Red Service
Required light is illuminated.Audible alarm may be sounding.
C. (DeVilbiss LT Series) Unit operates, the Power light is on
when the power switch is On. Red Service Required light is
illuminated.Audible alarm may be sounding.
D. (DeVilbiss 3 & 5 Liter Series) Limited flow of oxygen,audible alarm is
pulsing and Service Required light is flashing.
E. If any other problems occur with your oxygen concentrator.
Possible Cause
1. Power cord not properly inserted into wall outlet.
2. No power at wall outlet.
3. Oxygen concentrator circuit breaker activated
(only 115 volt units).
1. Air filter is blocked.
2. Exhaust is blocked.
1. Air filter is blocked.
2. Exhaust is blocked.
1. Blocked or defective cannula, catheter, face mask,
or oxygen tubing.
2. Blocked or defective humidifier bottle.
3. Flow meter set too low.
Remedy
1. Check power cord connection at the wall outlet. On 230 volt units,
also check the back of the unit.
2. Check your home circuit breaker and reset if necessary. Use a
different wall outlet if the situation occurs again.
3. Press the concentrator circuit breaker reset button located below
the power switch (only 115 volt units). Use a different wall outlet if
the situation occurs again.
If the above remedies do not work,contact your
Sunrise Medical provider.
1. Check the air filter. If the filter is dirty, wash it following the
cleaning instructions on page 18.
2. Check the exhaust area;make sure there is nothing restricting the
unit exhaust.
If the above remedies do not work,contact your
Sunrise Medical provider.
1. Check the air filter. If the filter is dirty, wash it following the
cleaning instructions on page 18.
2. Check the exhaust area;make sure there is nothing restricting
the unit exhaust.
If the above remedies do not work,contact your
Sunrise Medical provider.
1 Detach cannula,catheter, or face mask.If proper flow is restored,
clean or replace if necessary. Disconnect the oxygen tubing at the
oxygen outlet. If proper flow is restored, check oxygen tubing for
obstructions or kinks. Replace if necessary.
2. Detach the humidifier from the oxygen outlet. If proper flow is
obtained, clean or replace humidifier.
3. Set flow meter to prescribed flow rate.
If the above remedies do not work,contact your
Sunrise Medical provider.
1. Turn your unit Off. Switch to your reserve oxygen system,and
contact your Sunrise Medical provider immediately.
MINOR TROUBLESHOOTING
The following troubleshooting chart will help you analyze and correct minor oxygen concentrator malfunctions. If the suggested procedures do not help, switch to your reserve oxygen sys-
tem and call your Sunrise Medical homecare provider. Do not attempt any other maintenance.
WARNINGTo avoid electric shock hazard, do not remove the cabinet.The cabinet should only be removed by a qualified Sunrise Medical homecare technician.
MINOR TROUBLESHOOTING CHART
Visual Indicator Lights
A. Both the green Normal Oxygen and the yellow Low Oxygen lights
are either on or off.
B. Yellow Low Oxygen light is on or the yellow Low Oxygen light is on
and the intermittent audible signal is sounding.
Possible Cause
1. OSD malfunction.
1. Flow meter is not properly set.
2. Air filter is blocked.
3. Exhaust is blocked.
Remedy
1. Contact your Sunrise Medical provider.
1. Ensure the flow meter is properly set to the prescribed number.
2. Check the air filter. If the filter is dirty, wash it following the
cleaning instructions on page 18.
3. Check the exhaust area;make sure there is nothing restricting the
unit exhaust. If the above remedies do not work,contact your
Sunrise Medical provider.