ResMed S9 VPAP Tx Sleep Apnea Machine User Manual


 
28
Sound:
DECLARED
DUAL-NUMBER
NOISE
EMISSION
VALUES in
accordance with
ISO 4871:1996
Sound pressure level (CPAP mode)
with SlimLine air tubing: 26 dBA with uncertainty of 2 dBA as measured
according to ISO17510-1:2007
with Standard air tubing: 27 dBA with uncertainty of 2 dBA as measured
according to ISO17510-1:2007
with either SlimLine or
Standard air tubing and H5i:
28 dBA with uncertainty of 2 dBA as measured
according to ISO17510-1:2007
Sound power level (CPAP mode)
with SlimLine air tubing: 34 dBA with uncertainty of 2 dBA as measured
according to ISO17510-1:2007
with Standard air tubing: 35 dBA with uncertainty of 2 dBA as measured
according to ISO17510-1:2007
with either SlimLine or
Standard air tubing and H5i:
36 dBA with uncertainty of 2 dBA as measured
according to ISO17510-1:2007
Alarm volume
settings
Low (nominal 56 dBA), Medium (nominal 68 dBA), High (nominal 80 dBA)
Physical Dimensions (L x W x H): 6.0” x 6.8” x 3.4” (153 mm x 172 mm x 86 mm)
Weight: 2.30 lbs (1.045 kg)
Housing construction: Flame retardant engineering thermoplastic
Air outlet: 22 mm conical air outlet (complies with ISO 5356-1:2004)
Air filter Standard: Polyester non-woven fiber
Hypoallergenic: Acrylic and polypropylene fibers in a polypropylene carrier
Supplemental
oxygen
Recommended maximum supplemental oxygen: 15 L/min (CPAP, S, T, ST, ASV and
PAC modes), 4 L/min (AutoSet, VAuto, iVAPS and ASVAuto modes)
Pneumatic flow path
9
2
5
4
6
7
3
1
8
3
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. H5i
7. Alarm module
8. Device
9. Inlet filter