ResMed S9 VPAP Tx Sleep Apnea Machine User Manual


 
6
Setting up the S9 VPAP Tx
WARNING
Make sure the power cord and plug are in good condition and the equipment is not damaged.
Keep the power cord away from hot surfaces.
Do not leave long lengths of air tubing around the top of the patient’s bed. It could twist around
the patient’s head or neck while they are sleeping.
Do not use electrically conductive or antistatic air tubings.
Do not use the air tubing if there are any visible signs of damage.
Only ResMed products are designed to be connected to the module connector port. Connecting
other devices could damage the device.
CAUTION
Incorrect system setup may result in incorrect mask pressure reading. Ensure the system is
correctly set up.
Be careful not to place the device where it can be bumped or where someone is likely to trip
over the power cord.
The H5i should only be connected or disconnected when the water tub is empty.
Always place the H5i on a level surface below the level of the user to prevent the mask and
tubing from filling with water.
The S9 VPAP Tx is compatible with the integrated H5i heated humidifier. For further information on using
this humidifier refer to the H5i user guide.
1. Align the H5i with the S9 VPAP Tx and push them together until they click into place.
2. Connect the DC plug of the power supply unit to the rear of the S9 VPAP Tx.
3. Connect the power cord to the power supply unit.
4. Plug the USB Module into the Module/Adaptor port at the rear of the S9 VPAP Tx.
Module/Adaptor Port
To Tx Link
5
8
1
6
3
4
2
7
Info menu button
LCD screen
Setup menu button
Push dial
Start/Stop
button
Alarm and
therapy LEDs
Alarm mute
button
S9 VPAP Tx basics
Key Function
Start/Stop button
Starts or stops treatment.
Power Save mode—hold for three seconds.
Info menu button
Allows you to view the device service information or to exit from
the menu.
Setup menu button Allows you to make changes to settings or to exit from the menu.
Push dial
Turning the dial allows you to scroll through the menu and change
settings. Pushing the dial allows you to enter into a menu and
confirm your choice.
Alarm mute button
Press once to mute alarms. Press a second time to un-mute. If
the problem is still present, the alarm will sound again after two
minutes.
LCD screen
Displays the menus, treatment screens and reminders.
Backlight—when treatment is being delivered, the backlight
(including the Start/Stop button) automatically turns off after
30seconds, otherwise it turns off after 3 minutes.
Alarm LED Yellow LED—flashes during an alarm.
Therapy LED
Blue LED—always on during therapy (if enabled in the Options
menu).