Clinical setup menu parameters
*
Parameter Modes Range Description
CPAP
AutoSet
VAuto
S
ST
T
PAC
iVAPS
ASV
ASVAuto
Settings
Circuit
Mask type
9 9 9 9 9 9 9 9 9 9
Full Face / Nasal
/ Pillows / Nasal
Ultra / Pediatric
Default: Full Face
Selects the type of mask used
by the patient.
Tube type
9 9 9 9 9 9 9 9 9 9
SlimLine /
Standard / 3 m
Default: Standard
Shows the type of air tubing
used by the patient.
Not displayed if ClimateLine or
ClimateLine
MAX
is connected
AB filter
9 9 9 9 9 9 9 9 9 9
No / Yes
Default: No
Enables or disables antibacterial
filter.
Not displayed if H5i is
connected.
Ext.
humidifier
9 9 9 9 9 9 9 9 9 9
No / Yes
Default: No
Enables or disables an external
humidifier.
Therapy
Mode
9 9 9 9 9 9 9 9 9 9
CPAP / S / ST /
T / PAC / iVAPS /
ASVAuto / ASV /
AustoSet / VAuto
Default: CPAP
Sets the therapy mode available
on the device.
Set pressure
9
4–20 cm H
2
O
Default:
4 cm H
2
O
Sets the fixed treatment
pressure.
EPR Level
9 9
Off / 1 / 2 / 3
Default: 3
Sets the EPR value.
Height
9
44-100 inches
(110–250 cm)
Default: 70 inches
(175cm)
Body height needed for the
dead space determination.
Target Pt
Rate
9
8–30 BPM
Default: 15 BPM
The rate input to the iVAPS
algorithm. This should be set at
the patient’s actual respiratory
rate.
Target Va
9
1–30 L/min
Default: 5.2 L/min
Used to determine the amount
of pressure support required by
the iVAPS algorithm.
IPAP
9 9 9 9
4–30 cm H
2
O
Default:
10 cm H
2
O
Sets the pressure which will be
delivered to the patient when
the device is triggered into
inspiration.
*Default settings may differ from country to country