ResMed S9 VPAP Tx Sleep Apnea Machine User Manual


 
20
Managing Climate Control
Designed to be ideal for most patients, Climate Control Auto enables the automatic delivery of a constant
value of absolute humidity while protecting against rainout.
To allow for increased flexibility, Climate Control can be turned to Manual in the Setup menu. Setting
Climate Control to Manual disables the automatic control of humidity and allows humidity and
temperature levels to be set independently. However, rainout protection is not provided when Climate
Control is set to Manual.
Mode
Humidity Temperature
Setting range Default settings Setting range Default settings
Climate Control – Auto
Climate Control
Constant
absolute humidity
(depending on
temperature
setting)
Off*, 60ºF–86ºF
(16ºC–30ºC)
80ºF (27ºC)
Climate Control –
Manual
Humidity level Heated tube
Off–6.0
(0.5 increments)
3 Off, 60ºF–86ºF
(16ºC–30ºC)
80ºF (27ºC)
*
When the temperature setting is set to Off the tube will not heat the air, nor will the humidifier heat the water to
add humidity to the air.