ResMed C16389 Respiratory Product User Manual


 
28
!
CAUTION
Do not attempt to open the VPAP Malibu. There are no user serviceable
parts inside. Repairs and internal servicing should only be performed by an
authorized service agent.
Replacing the Air Filter
Inspect the air filter every month to check if it is blocked by dirt or contains holes.
With normal use of a VPAP Malibu unit, the air filter needs to be replaced every
six months (or more often if your unit is in a dusty environment). To replace the
air filter:
Note: For details of how to check and replace the hypoallergenic filter (if used)
see the Hypoallergenic Air Filter User Instructions.
1. Remove the air filter cover at the back of the VPAP Malibu.
2. Remove and discard the old air filter.
3. Insert a new filter with the blue tinted side facing out.
4. Replace the air filter cover.
!
WARNING
Do not wash the air filter. The air filter is not washable or reusable.
Replacing the Antibacterial Filter
You should regularly inspect and replace the antibacterial filter, as outlined in the
Antibacterial Filter User Instructions.
Note: ResMed recommends that you use the Pall BB50TE filter.
Servicing
This product (VPAP Malibu) should be inspected by an authorized ResMed
Service Center 5 years from the date of manufacture. Prior to this, the device is
intended to provide safe and reliable operation provided that it is operated and
maintained in accordance with the instructions provided by ResMed. Applicable
ResMed warranty details are provided with the device at the time of original
supply. Of course, as with all electrical devices, if any irregularity becomes
apparent, you should exercise caution and have the device inspected by an
Authorized ResMed Service Center.
Air
filter
cover
268106r1.book Page 28 Thursday, November 30, 2006 2:35 PM