25Operating Instructions
English
Table 1: Definitions of Mask-Fit Star Rating
Using DC Power to Run the VPAP Malibu
The VPAP Malibu can be run on DC in the case of power failure. ResMed
supplies separately a converter that allows a 12V or 24V battery to be connected
to the DC input of the VPAP Malibu. When using DC, power is turned on and off
via the switch on the connector to the power source, not the power switch at the
back of the VPAP Malibu. The DC converter is a separate accessory that can be
purchased through your ResMed distributor.
The humidifier heater will be automatically disabled when the VPAP Malibu is
operating from the DC supply.
Power consumption of the VPAP Malibu varies with treatment settings.
Adequate battery type and size is required before operation on DC, and it is
recommended that you contact ResMed’s Technical Services division or
ResMed's website for information on your specific application.
!
CAUTION
• The VPAP Malibu uses 24V DC power. Only connect the VPAP Malibu to
DC power using the ResMed DC-24/30 Converter specified for this
application.
• Do not connect both AC and DC power cords to the VPAP Malibu at the
same time.
Helpful Hints
Starting Out
Mouth Leaks
When using a nasal mask, try to keep your mouth closed during treatment. Air
leaks from your mouth can decrease the effectiveness of your treatment. If
mouth leaks are a problem, a full face mask or chin strap may help. Contact your
clinician or equipment supplier for further details.
Star rating Definition
***** Excellent
****- Very good
***- - Good
**- - - Adjust mask
*- - - - Adjust mask
HIGH LEAK Adjust mask
268106r1.book Page 25 Thursday, November 30, 2006 2:35 PM