VI. BATTERIES AND CHARGING
BATTERIES AND CHARGING
The Quantum 610 uses two long-lasting, 12-volt, deep-cycle batteries. These batteries are sealed and maintenance free.
Since they are sealed, there is no need to check the electrolyte (fluid) level. Deep-cycle batteries are designed to handle a
longer and deeper discharge. Though they are similar in appearance to automotive batteries, they are not interchangeable.
Automotive batteries are not designed to handle a long, deep discharge, and also are unsafe for use in power chairs.
WARNING! Battery posts, terminals, and related accessories contain lead and lead compounds.
Wash hands after handling.
WARNING! Prevent injury and/or equipment damage! Do not use batteries that exceed the
recommended type and amp-hour capacity. Do not use batteries with different amp-hour capacities.
Refer to specifications table in this manual and in the manual suppled with the battery charger
for recommended type and capacities.
Charging the Batteries
The battery charger is essential in providing long life for your power chair batteries. It is designed to optimize your power
chair’s performance by charging the batteries safely, quickly, and easily. The battery charging system consists of the battery
charger, the charger circuit fuse, and the ammeter (optional onboard chargers only). The ammeter indicates the rate of
charge necessary to fully recharge the batteries. It is also a good indication of whether or not the charger is working. The
ammeter and the charger are only functional when the charger power cord is plugged into an electrical outlet. The charger
circuit has an ATO fuse that protects the ammeter. See figure 7.
PROHIBITED! Removal of grounding prong can create electrical hazard. If necessary, properly
install an approved 3-pronged adapter to an electrical outlet having 2-pronged plug access. Failure
to heed could result in personal injury and/or property damage.
PROHIBITED! Never use an extension cord to plug in your battery charger. Plug the charger directly
into a properly wired standard electrical outlet.
WARNING! You must recharge your power chair’s batteries with the supplied off-board charger or
optional onboard battery charger. Do not use an automotive-type battery charger.
WARNING! Prevent injury and/or equipment damage! Read the battery charging instructions in
this manual and in the manual supplied with the battery charger before charging the batteries.
WARNING! Prevent injury and/or equipment damage! Do not expose the off-board battery charger
to rain or other sources of moisture unless it has been tested for outdoor use. Refer to the manual
supplied with the battery charger for more information.
WARNING! Prevent injury and/or equipment damage! Explosive gases may be generated while
charging the batteries. Keep the power chair and battery charger away from sources of ignition
such as flames or sparks and provide adequate ventilation when charging the batteries.
WARNING! Prevent injury and/or equipment damage! Inspect the battery charger, wiring, and
connectors for damage before each use. Contact your Quantum Rehab Specialist if damage is
found.
WARNING! Prevent injury and/or equipment damage! Do not attempt to open the battery charger
case. If the battery charger does not appear to be working correctly, contact your Quantum
Rehab Specialist.
WARNING! Prevent injury and/or equipment damage! If the off-board battery charger is equipped
with cooling slots, then do not attempt to inset objects through these slots.
WARNING! Prevent injury and/or equipment damage! Do not allowed unsupervised children to
play near the power chair while the batteries are charging.