Omega 27 Thermometer User Manual


 
18
6. Select the resolution desired (0.1°/1° key).
NOTE: With 1° resolution, one less key stroke will be required to
enter calibrator output temperatures via the keypad.
7. The unit accepts numeric inputs by first changing the keypad to its num-
ber-entry mode (NUM) by depressing the CHANGE/ENTER key once.
The numeric display will flash, the OPR annunciator will turn off, and the
NUM annunciator will turn on. Key in desired temperature value, most sig-
nificant digit first. After the desired temperature value is keyed in, hit
CHANGE//ENTER to return to OPR mode.
7. To enter a value of 212° with 0.1° resolution, the following keystrokes are
used:
CHANGE/ENTER, 2, 1, 2, 0, CHANGE/ENTER
NOTES:
Invalid numeric entries (eg. a temperature beyond the span of
the selected TC type) will cause a corresponding error code to
be displayed.
To enter negative (ie. minus) temperatures, hit the –0 key first
when in the NUM mode. The first hit on this key enters the neg-
ative sign. Subsequent hits enter zeros.
To correct a numeric entry while in the NUM mode, hit the
CHANGE/ENTER key twice. This will return number entry to
the beginning of a new number.
To return the calibrator output to 0° while in the OPR mode, hit
the CLEAR key.
The unit can be used as a convenient °F to °C or °C to °F cal-
culator.
For example, to convert 77°F to a Celsius reading, go to the CALIB OPR
mode, enter the reading to be converted, and change scale.
eg. °F
eg. CHANGE/ENTER
eg. 77
eg. CHANGE/ENTER
eg. °C (Display reads 25°C)
7. CALIBRATOR MODE OPERATION (RTD, Thermistor & Ohms)
Set-Up:
1. Connect measurement equipment under test (eg. RTD thermometer,
ohmmeter, etc.) to the instrument as shown in Figures 8, 9, or 10.