Omega Speaker Systems OS550-BB Series Thermometer User Manual


 
7-3
Specifications
7
1 mV/degree C or F Analog Output:
Accuracy: ± 2 mV referenced to the temperature display
0-5 Vdc Analog Output:
Accuracy: ± 0.25% of full scale (Referenced to display reading)
Scaling: Fixed (scaled to match full temperature range of
model selected)
Minimum Load: 600 ohms
4-20 mA Analog Output:
Accuracy: ± 0.25% of full scale (Referenced to display reading)
Scaling: Fixed (scaled to match full temperature range of
model selected)
Maximum Load: 900 ohms @ 24 Vdc