Omega Speaker Systems OS550-BB Series Thermometer User Manual


 
Using the Infrared Thermometer/Transmitter
3-18
3
Figure 3-10. Serial Printer Hookup
3. With power applied to the unit, press and hold the key
until the Print Data Display Mode (PRN) appears.
4. Press the key to increment the printing interval.
Press the key to decrement the printing interval. The printing
interval (from 1 to 1999 seconds) is the time between data points.
The default value is 2 seconds.
5. Press the key to start sending data. The icon
appears on the display.
To stop sending data, press the key again, and the
icon disappears.
RS-232
Digital
Cable
25-Pin 'D'
Connector
6-Pin
Phone Jack
To the
Printer
To the
Pyrometer
Serial Printer
NOTE
Bottom hole is
the RS-232 jack
°F
PRN
NOTE