Lifescan Ultra 2 Blood Glucose Meter User Manual


 
42 Customer Service 1 800 227-8862
Control solution testing
OneTouch
®
Ultra
®
Control Solution contains
a known amount of glucose and is used to
check that the meter and the test strips are
working properly.
Do a control solution test:
to practice the test process instead of
using blood,
once a week,
whenever you open a new vial of test strips,
if you suspect the meter or test strips are not working properly,
if you have had repeated unexpected blood glucose results (as described
on pages 28 and 29), or
if you drop or damage the meter.
NOTE:
• Use only OneTouch
®
Ultra
®
Control Solution with your
OneTouch
®
Ultra
®
2 Meter.
• Control solution tests must be done at room temperature (68–77°F).
Make sure your meter, test strips, and control solution are at room
temperature before testing.
ƽ CAUTION: Do not swallow
control solution; it is not for
human consumption.
Do not apply control solution
to the skin or eyes as it may
cause irritation.
DCO_06344501B_OU2_OB_USe.qxd 12/21/05 8:03 AM Page 42