Checking the System with
PocketScan™ Control Solution
PocketScan Control Solution is used to
check that the meter and test strips are
working together as a system and that
you are performing the test correctly. It is
very important that you do this simple
check routinely to make sure you get
accurate results.
PocketScan Control Solution contains a
known amount of glucose that reacts with
PocketScan™ Test Strips. Compare your
control solution test results with the expect-
ed range printed on the test strip vial label.
Control solution should be used to:
Practice the test procedure.
Make sure your meter and test strips are
working together properly.
Ensure you are performing the test cor-
rectly.