Leica FSC Microscope & Magnifier User Manual


 
16
7. Assembly
Coaxial incident light
Remove the cover plates (9.1) to the rear left
and right on the macroscope carriers.
In their place, screw on the coaxial
illuminator (11.1)
Screw the light guide with ø 9 mm (Fig. 12)
onto the cold light illuminator (8.1).
Place the cold light source next to the stand
or onto the shelf that is to be screwed onto
the base of the comparison macroscope (7.1).
Insert the polarization filter cube (Fig. 10).
To do so, attach the filter cube onto the side
and push it into the holder (11.2).
Fig. 9 Fastening the coaxial illuminator
1 Fastening screws for the cover plate
Fig. 11 Coaxial illuminator
1 Fastening screws for the coaxial illuminator
2 Filter cube inserted
3 Receptacle for light guide
Fig. 10 Polarizing
filter cube
1 From the rear
2 From the front
1
1
2
3
1
2
Insert the two light guide ends into the left and
right receptacles on the coaxial illuminator
(11.3).
Note:
If, after installation, point-like objects are ap-
pear in the image, the light guide must be pulled
1 to 2 mm out of the coaxial illuminator. The
point-like objects are caused by the fibers of the
light guide.