5. Adjust the objective centering screws
until the noticeable point of the sample is
midway between the furthest position and
the center of the crosslines. This moves the
objective center closer to the stage center.
6. Move the specimen (not the objective
centering screws!) so that the noticeable
point is at the center of the crosslines once
again.
7. Check to make sure that the noticeable
point stays at the center of the crosslines
when the stage is rotated. If not, repeat the
centering process.
Each objective must be centered sepa-
rately. Also, there is a magnetic storage
location for the objective centering tools in the
back of the Analyzer or Analyzer/Bertrand Lens
modules.
Objective Centration (cont.)