Leica B523 Microscope & Magnifier User Manual


 
15
Cover the microscope with the red hood
6 Troubleshooting
6.1 There is no light
Microscope is not turned on (see Section 2.1)
Light intensity set too low (see 2.1 on page 1)
Do you see any light in the lamp chamber? If not, the halogen bulb may need
to be changed - contact MIU
6.2 Light is too dark
Check:
whether the polarizer or other filters are on the light path (Section 2.9 on page
8)
that aperture diaphragm is not too closed (Section 2.8 on page 6)
6.3 Problems with focus or picture clarity
Check that:
no filters are on the light pathway
Koehler illumination is correctly done (Section 2.8 on page 6)
objective is not dirty
oculars are not dirty
oculars are correctly adjusted (Section 2.6)
If image quality does not satisfy you, you should take your images with AxioPlan
microscope, which has better optics and camera
6.4 View is partial
Check that:
none of the filters are partially on the light path
the polarizer is not partially on the light path
the objective is correctly in place
the field diaphragm is not too closed
6.5 Colors have red/yellow tint
Check the white balance
For other color errors, check the color filters (Section 2.10)
6.6 Ocular view is not at the same focus as the camera view
Check ocular adjustment (Section 2.6)