Hanna Instruments HI 4222 Thermometer User Manual


 
42
Unrecognized buffer. Please check the buffer or the buffer list (for Semiautomatic and Automatic buffer entry
type): this message appears if the current buffer value is not close to any of the buffers from the buffer list/group.
Check if the current buffer is present in the buffer list or the appropriate buffer group was selected.
The current buffer was already calibrated. Press <Accept> to recalibrate in this point, or change
the buffer: this message appears when calibration is performed in one of the previous calibrated buffers.
Follow displayed instructions.
pp
pp
p
H MEASUREMENTSH MEASUREMENTS
H MEASUREMENTSH MEASUREMENTS
H MEASUREMENTS
Make sure the instrument has been calibrated before taking pH measurements.
DIRECT MEASUREMENT
To measure the pH of a sample using the Direct reading mode:
Press
and then to select
pH Measure
mode
(for the selected channel – HI 4222 only).
Select the Direct reading mode (see pH Setup for details, page 20).
Submerge the electrode tip and the temperature probe
approximately 4 cm (1½”) into the sample to be tested.
Allow time for the electrode to stabilize.
The measured pH value will be displayed on the LCD, together
with a short GLP information or “Not Calibrated” message if
no pH calibration was performed.
Note: If the reading is out of range, “-----” will be displayed
on the LCD.
DIRECT / AUTOHOLD MEASUREMENT
To measure pH of a sample using the Direct/AutoHold reading mode:
Press
and then to select
pH Measure
mode
(for the selected channel – HI 4222 only).
Select the Direct/AutoHold reading mode (see pH Setup for
details, page 20).
Submerge the electrode tip and the temperature probe
approximately 4 cm (1½”) into the sample to be tested.
Allow time for the electrode to stabilize.
The measured pH value will be displayed on the LCD. If pressing
, the “AutoHold” indicator will start blinking on the LCD
until the stability criterion is reached. The pH value will be
frozen on the LCD, along with “AutoHold” indicator.